KEMZ 801

EM Clamp for IEC/EN 61000-4-6

KEMZ 801Azoom
  • As specified in IEC/EN 61000-4-6
  • Very efficient coupling
  • Can be used on almost any cable
  • Ruggedly designed
 

IEC 61000-4-6 (EN 61000-4-6) defines three basic types of transducer used for the injection of test signals into the equipment under test (EUT). The basic requirement is to inject a known level of RF signal onto the cable of the EUT at each test frequency and to determine whether the EUT continues to function correctly. In order to do this, it is necessary to decouple the auxiliary equipment (AE) from the test signal. If this is not achieved it is difficult to know whether any fault is due to a failure of the EUT or the AE. Whilst the coupling/decoupling network (CDN) is one of the best ways to achieve this test and indeed it is mandatory to use them under some circumstances, it is not always possible to use a CDN.

Parameter Value
Frequency range

10 kHz to 1000 MHz

Nominal impedance

50 Ω

Connector

N-type female

Maximum input level

0.01 to 100 MHz: 100 W for 15 min
100 to 230 MHz: 100 W for 3 min
230 to 1000 MHz: 50 W for 3 min

Maximum correction factor

0.15 to 400 MHz: 4 dB (50 Ω system), 13.5 dB (150 Ω system)

  • Related Products

    Description

     
  • KEMA 801

    KEMA 801

    The KEMA 801 is recommended as an additional decoupling network (ferrite tube clamp) for immunity test according to IEC/EN 61000-4-6 when using the clamp injection method.

Datasheet

Gallery

 
 
 
 

Imprint

AMETEK CTS GmbH
Sternenhofstr. 15
4153 Reinach, Switzerland
Tel.+ 41 61 204 41 00
marketing.cts@ametek.com

Managing Directors:
Manfred A. Bergsch
Astrid Wälchli Schreiber
VAT-ID: CHE-179.452.414
Register of companies: CH-280.4.015.069-3


 
 
By continuing to use the site, you agree to our Privacy and Cookie Policy.